Method for the determination of correct calibration pulses in PEA measurements
编号:11 访问权限:仅限参会人 更新:2022-08-29 10:48:25 浏览:161次 口头报告

报告开始:暂无开始时间(Asia/Shanghai)

报告时间:暂无持续时间

所在会场:[暂无会议] [暂无会议段]

视频 无权播放 演示文件

提示:该报告下的文件权限为仅限参会人,您尚未登录,暂时无法查看。

摘要
IEEE ICHVE 2022 / 1-PAGE ABSTRACT
Method for the determination of correct calibration pulses in PEA measurements
P. Seri, D. Mariani, D. Fabiani
Department of Electrical, Electronic and Information Engineering (DEI)
daniele.mariani8@unibo.it

Purpose/Aim
The Pulsed Electro-Acoustic (PEA) method is being increasingly adopted by institutions and industries in the development of dielectrics for HVDC applications. A correct PEA measurement always requires postprocessing of raw data coming from the experimental setup (the so-called calibration of the signal). One fundamental step required to correctly calibrate the PEA signal is the identification of a time instant at which no space charge is being accumulated in the specimen. This can be done in different ways, with different levels of reliability. This work proposes an alternative method, which minimizes the possibility of an incorrect evaluation regarding space charge accumulation at the very beginning of polarization in a material.
Experimental/Modeling methods
Experimental setup is based on the introduction of a floating mass in the thickness of the material to be tested. This might be done fusing multiple layers of insulating materials together, and embedding a floating electrode. Probing of the potential of the metallic interfacial electrode can be done both when the specimen is under an AC supply, and also in DC, during polarization and depolarization phases. The evolution of the detected potential over time, when no space charge is being injected, must be proportional to the potential applied to the test sample, as the system acts as a voltage divider. Discrepancies of such proportionality factor over time should indicate that space charge accumulation is being injected in the specimen.
Results/discussion
Tests were run poling specimens both in AC and in DC. Analysing the time evolution of the proportionality factor between the applied voltage and the floating mass potential, there is clear indication that an amount of space charge is being injected, in DC, after a certain time, depending on the kind of insulation and testing conditions. Hence, the calibration pulse should be chosen from acquisitions within such time.
Conclusions
The calibration process of PEA measurements is prone to uncertainties, since a signal supposedly free of charge accumulation should be selected. The proposed method is a viable option to ensure that is indeed the case, with much less uncertainties than more traditional methods.
 
关键词
Dielectrics,space charge,HVDC,pulsed electroacoustic methods
报告人
Daniele Mariani
Eng. University of Bologna

稿件作者
Simone Vincenzo Suraci DEI - University of Bologna
Paolo Seri University of Bologna
Daniele Mariani University of Bologna
Fabiani Davide DEI - University of Bologna
发表评论
验证码 看不清楚,更换一张
全部评论
重要日期
  • 会议日期

    09月25日

    2022

    09月29日

    2022

  • 08月15日 2022

    提前注册日期

  • 09月10日 2022

    报告提交截止日期

  • 11月10日 2022

    注册截止日期

  • 11月30日 2022

    初稿截稿日期

  • 11月30日 2022

    终稿截稿日期

主办单位
IEEE DEIS
承办单位
Chongqing University
移动端
在手机上打开
小程序
打开微信小程序
客服
扫码或点此咨询