Analysis of Spectral Calibration Correction Based on Mie Scattering
编号:82
访问权限:仅限参会人
更新:2024-10-23 10:37:26
浏览:168次
口头报告
摘要
Oblique light incidence on a diffuse transmission plate significantly impacts the results in the spectral calibration process. Analysis of the spectrometer's diffraction grating reveals that oblique angles cause uneven energy distribution at the entrance pupil, leading to spectral shifts. This study uses Mie scattering theory to model the diffuse transmission plate and explains the spectral shift, particularly in the infrared region where the plate's micron-scale particles have a pronounced effect. For a 1064nm laser incident at 45° on a 50-micron particle plate, the calculated spectral shifts at 45°, 55°, and 65° are 0.077nm, 0.108nm, and 0.139nm, respectively. Experimental Gaussian-fitted results show actual shifts of 0.044nm, 0.052nm, and 0.085nm. The paper concludes with an analysis of the differences between experimental and theoretical results.
关键词
Spectral calibration, Mie scattering, Grating system
稿件作者
YuQi
University of Chinese Academy of Sciences
HuangYu
Chinese Academy of Sciences;Changchun Institute of Optics; Fine Mechanics and Physics
发表评论